2.1 Static Performance Parameters

Measurement is the process of comparing an unknown physical quantity with a predefined standard unit to determine its numerical value. When an input is applied to an instrument, the output undergoes an initial transient period before settling to a constant steady-state period.

Definition

Static Performance Parameters

Static performance parameters (static characteristics) describe the behavior and quality of a measuring instrument when measuring quantities that are constant or vary very slowly with time (steady-state operation).

Accuracy

Accuracy specifies the closeness with which an instrument reading approaches the true value (T.V.) of the measured quantity. Deviation between the measured value (M.V.) and the true value indicates how accurately a measurement has been made. Accuracy is specified in three standard ways (pages="notes"):

  1. Point Accuracy: Accuracy guaranteed only at one specific point on the instrument's scale.
  2. Percentage of Scale Range (Full-Scale Accuracy): For uniform-scale instruments, maximum error is stated as a percentage of the full-scale reading. Notice that measuring small quantities near the bottom of the scale produces a much larger percentage error relative to the true value.
  3. Percentage of True Value: Accuracy expressed directly as a percentage of the actual quantity being measured—the most informative specification.

Precision (Repeatability)

Precision specifies the repeatability or closeness of agreement among a set of independent measurements of the same quantity made with the same instrument under identical conditions.

AspectAccuracyPrecision
MeaningCloseness of measured value to the true valueCloseness/repeatability among repeated readings
Example ()A single reading of is accurateRepeated readings of (even if offset from ) are highly precise
DependencyAccuracy requires precisionPrecision does not guarantee accuracy (a defective/uncalibrated meter can be precise yet inaccurate)

Sensitivity, Resolution, and Linearity

Definition

Sensitivity, Resolution, and Linearity

When the calibration curve is a straight line, sensitivity is constant across the operating range.


2.2 Dynamic Performance Parameters

When the measured input changes rapidly, an instrument does not reach its steady-state reading instantaneously. Dynamic performance parameters characterize the transient response between the instant an input changes and the time the output settles to its final value.

Definition

Response Time, Frequency Response, and Bandwidth

where and are the lower and upper cutoff frequencies.


2.3 Error in Measurement

No physical measurement is completely free of error. Error is defined as the deviation of the measured value () from the true value () of the quantity:

Classification of Measurement ErrorsText
                        Measurement Errors
                                |
        +-----------------------+-----------------------+
        |                       |                       |
   Gross Errors         Systematic Errors         Random Errors
                                |
                    +-----------+-----------+
                    |                       |
           Instrumental Errors     Environmental Errors
Types of Measurement Errors, Causes, and Corrective Measures
Error TypeDefinition & Common CausesCorrective Measures
Gross ErrorsHuman mistakes by the observer: misreading scales, wrong circuit connections, recording or calculation errors.Take readings and perform calculations carefully; take at least independent readings (preferably by different observers).
Systematic — InstrumentalShortcomings of the instrument itself: poor scale calibration, defective components, worn bearings, or improper spring tension.Use calibrated standard instruments, apply calibration correction factors, and recalibrate periodically.
Systematic — EnvironmentalExternal surroundings: variations in temperature, pressure, humidity, or stray electrostatic and magnetic fields.Use protective/sealed casings against humidity, perform measurements in temperature-controlled rooms, and apply magnetic/electrostatic shielding.
Random ErrorsUnpredictable variations from unknown causes that remain even after all gross and systematic errors are removed.Apply statistical analysis (take multiple repeated readings and compute the arithmetic mean and standard deviation).

2.4 Statistical Analysis of Error in Measurement

Statistical methods are applied after minimizing gross and systematic errors to estimate the most probable true value and quantify random scatter (pages="notes").

Let be repeated readings of a quantity:

  1. Arithmetic Mean (): The most probable value of the measured variable:
  1. Deviation from Mean (): The departure of reading from the arithmetic mean:
  1. Average Deviation (): An indicator of instrument precision, defined as the mean of the absolute deviations:
  1. Standard Deviation ( or ): The root-mean-square deviation:
  1. Variance (): The square of the standard deviation:

Gaussian Probability Curve and Probable Error

When a large number of repeated readings affected only by random errors are plotted as a histogram at finer and finer intervals, the contour approaches a symmetric bell-shaped Gaussian (normal) probability curve:

A sharper, narrower curve indicates higher precision and stronger confidence that lies close to the true value. The probable error () defines the limits that encompass of all observations (an even chance that any single reading falls within ):

Fraction of Total Observations Within Gaussian Deviation Limits
Deviation LimitsFraction of Total Area IncludedPercentage of Readings
(Probable error )

2.5 Measurement of Resistance (Low, Medium and High)

Resistance values encountered in engineering span many orders of magnitude, requiring different measurement techniques for low, medium, and high resistances.

Classification of Resistance Ranges and Measurement Methods
CategoryResistance RangeStandard Measurement Methods
Low ResistanceAmmeter–Voltmeter method, Kelvin's double bridge, Potentiometer method
Medium ResistanceWheatstone bridge, Ammeter–Voltmeter method, Ohmmeter, Substitution method
High ResistanceLoss of Charge method, Megger (insulation tester), Direct deflection method

Low Resistance Measurement ()

  1. Ammeter–Voltmeter Method (pages="notes"): In the circuit where the voltmeter (resistance ) is connected directly across the unknown resistor , the voltmeter measures the true voltage , while the ammeter measures the total current :

Solving for the true resistance :

For accurate measurement without large correction, , making this configuration suitable for low resistances.

  1. Kelvin's Double Bridge Method (pages="added"): Standard Wheatstone bridges fail below because test-lead and contact resistances become comparable to the unknown . Kelvin's double bridge adds a second pair of ratio arms () across the yoke resistance connecting the unknown and standard . When , the effect of lead/yoke resistance cancels completely and balance gives:
  1. Potentiometer Method (pages="added"): The unknown low resistor is connected in series with a known standard resistor carrying a steady DC current . A precision potentiometer measures the potential drops across and across without drawing current, yielding:

Medium Resistance Measurement ()

The Wheatstone Bridge (analyzed in Section 2.6) is the primary laboratory method for medium resistances (pages="notes"). Other methods include the ammeter–voltmeter method, substitution method, and ohmmeters (pages="added").

High Resistance Measurement ()

Measuring insulation and high resistances poses two special difficulties (pages="added"): (i) currents are extremely small ( to ), and (ii) surface leakage currents across insulators can bypass the test specimen unless intercepted by a guard wire / guard ring connected to bypass the detector.

  1. Loss of Charge Method (pages="added"): A high-insulation capacitor is charged to an initial voltage and then discharged through the unknown high resistance for a measured time . The decaying voltage gives:
  1. Direct Deflection Method (pages="added"): A known DC supply drives a tiny current through in series with a sensitive galvanometer or microammeter; a guard ring shunts surface leakage straight back to the supply negative terminal so only volume current is measured.
  2. Megger (pages="added"): A portable high-voltage DC generator ( to ) paired with a crossed-coil ohmmeter movement used directly in the field to test cable and winding insulation resistance in to .

2.6 DC / AC Bridge (Wheatstone Bridge, Maxwell's Bridge, Schering Bridge)

Wheatstone Bridge (DC Bridge)

The Wheatstone bridge consists of four resistive arms (), a DC source , and a sensitive null detector connected between junctions and (pages="notes").

Wheatstone Bridge SchematicText
             a
            / \
         P /   \ Q
          /     \
         b--(D)--d
          \     /
         R \   / S (Rx)
            \ /
             c
     [DC Source V across a-c]

At null balance, the potential difference across the detector is zero (, so ), meaning , or . Since no current enters the detector:

Substituting into :

With the unknown resistance in arm 4, fixed ratio arms , and calibrated variable standard arm :

General AC Bridge Balance Condition

An AC bridge replaces the four resistors with complex impedances , , , and , excited by an AC source and monitored by an AC null detector (headphones, vibration galvanometer, or CRO) (pages="notes").

Theorem

Balance Conditions for an AC Bridge

For zero detector current (), the arm impedances must satisfy:

Separating into magnitude and phase yields two simultaneous balance conditions:

  1. Magnitude condition: (product of opposite impedance magnitudes are equal).
  2. Phase condition: (sum of opposite phase angles are equal).

Maxwell's Bridge (Inductance Measurement, )

Maxwell's inductance–capacitance bridge measures an unknown inductance (with series resistance ) in terms of a known standard capacitance (pages="notes"). Arm 1 contains , while arms 2 and 3 are pure resistances and :

At balance, :

Equating real and imaginary parts gives:

Practical Q-factor limits (pages="notes"): The coil quality factor is . Because , the phase balance requires , so .

Hence Maxwell's bridge is best suited for medium- coils ().

Schering Bridge (Capacitance Measurement)

The Schering bridge measures an unknown imperfect capacitor (modeled as series and loss resistance ) along with its dissipation factor (pages="notes"):

At balance, :

Equating real and imaginary parts:

The dissipation factor () of the capacitor is:


Worked Out Examples

Example 1

1

A voltmeter having a full-scale accuracy of over a range is used to measure (i) and (ii) . Determine the maximum percentage error in each reading.

Computing the absolute full-scale uncertainty:

Thus at the meter can read to , and at it can read to .

Percentage error at and :

  • For :
  • For :
Answer:

Maximum percentage error is at and at .

Example 2

2

Six independent voltmeter readings are recorded as . Compute (a) arithmetic mean, (b) average deviation, (c) standard deviation, (d) variance, and (e) probable error.

Arithmetic mean and individual deviations:

Deviations : (check: ).

Average deviation ():

Standard deviation (), variance (), and probable error (): Since , we use in the denominator:

Answer:

, , , , .

Example 3

3

The resistance is determined from , where current is measured with error and power with error. Find the maximum possible relative error in (both via worst-case bounds and first-order differential approximation).

Worst-case measured resistance: Resistance is maximized when is at its upper limit () and is at its lower limit ():

First-order error propagation:

Answer:

Maximum relative error in is (linear approximation) and (exact worst-case upper bound).

Example 4

4

A Maxwell's bridge operating at balances with , (parallel in arm 1), , and . Calculate the unknown coil's resistance , inductance , and quality factor .

Calculating and from Maxwell's balance equations:

Checking the coil's quality factor at :

Since , Maxwell's bridge is well suited for this coil.

Answer:

, , and .


Exercises

Exercises

Question 1

Differentiate between accuracy and precision. Define sensitivity, resolution, and linearity of an instrument.

Accuracy vs. precision:

  • Accuracy is how close a measured reading is to the true value of the quantity.
  • Precision is the degree of repeatability among multiple independent measurements of the same quantity. High precision does not guarantee accuracy if a systematic offset exists.

Sensitivity, resolution, and linearity:

  • Sensitivity: Ratio of change in output to change in input ().
  • Resolution: Smallest increment in input that produces a detectable output change (scale division for analog, LSB for digital).
  • Linearity: Closeness of the instrument's calibration curve to an ideal straight line (constant sensitivity across the range).
Answer:

Accuracy measures closeness to the true value, whereas precision measures repeatability across readings. Sensitivity is , resolution is the smallest detectable input step, and linearity measures conformity to .

Question 2

Classify errors in measurement into gross, systematic, and random errors, and state two corrective measures for each category.

Error categories and corrective measures:

  1. Gross Errors (human observation/connection/calculation mistakes): check wiring carefully and take at least 3 independent readings (preferably by different observers).
  2. Systematic Errors:
    • Instrumental (calibration drift, worn parts): use standard calibrated instruments and apply correction factors / recalibrate.
    • Environmental (temperature, humidity, stray fields): use sealed housings, temperature-controlled rooms, and electromagnetic shielding.
  3. Random Errors (unpredictable residual variations): take repeated readings and apply statistical analysis (arithmetic mean and standard deviation) after eliminating gross and systematic errors.
Answer:

Errors are classified as gross (careful reading + multiple observers), systematic—split into instrumental (recalibration + correction factors) and environmental (shielding + controlled environment)—and random (repeated trials + statistical analysis).

Question 3

A pressure transducer converts linearly into , measured on a analog voltmeter having equal divisions and a full-scale accuracy of . Find (a) the transducer sensitivity in , (b) the pressure resolution corresponding to scale division, and (c) the maximum percentage error when measuring .

Sensitivity and scale resolution:

  • Sensitivity: .
  • One voltmeter division , which corresponds to .

Maximum percentage error at : At , true output voltage . Full-scale uncertainty is , so:

Answer:

(a) , (b) per division, (c) .

Question 4

Five repeated ammeter readings are . Calculate the arithmetic mean, average deviation, variance, standard deviation, and probable error.

Mean and average deviation:

Deviations: .

Variance, standard deviation (), and probable error:

Answer:

, , , , .

Question 5

In a low-resistance measurement using the ammeter–voltmeter method (voltmeter connected directly across the resistor), the voltmeter () reads and the ammeter reads . Find the measured resistance , the true resistance , and the percentage error if voltmeter current is neglected.

Apparent (measured) resistance :

True resistance and percentage error: Using (or noting , so ):

Answer:

, true , percentage error .

Question 6

Explain the Loss of Charge method for high resistance measurement. A capacitor charged to discharges through an unknown insulation resistance so that its terminal voltage falls to in . Calculate .

Principle of Loss of Charge method: A known low-leakage capacitor is charged to voltage and connected across the unknown high resistance . As charge leaks through , the voltage decays exponentially as .

Computing :

Answer:

.

Question 7

Derive the balance equations for a Schering bridge. A Schering bridge balanced at has in parallel with in arm 1, in arm 2, and in arm 3. Find the unknown series capacitance , loss resistance , and dissipation factor .

Deriving Schering bridge balance: With , , , and :

Equating real and imaginary components gives and .

Substituting numerical values:

Answer:

, , and dissipation factor .